HAMILTON Calibration of Hamilton Syringes
HAMILTON SYRINGE SELECTION GUIDE
The calibration of Hamilton syringes is done by means of a gravimetric measurement. At 80% of the nominal syringe volumes, 5 to 10 measurements are done, depending on the syringe type and volume. Measurements at other positions than 80% of the nominal volume are not necessary, because the inner diameter of the syringe barrel is constant. Therefore, the volume changes in a linear way when the plunger is moved.
Proceeding
1. Aspiration and dispensation of an exact volume of deionized water.
2. Determination of the mass of the dispensed water.
3. Calculation of the volume of the dispensed water with mass and density.
4. Documentation of the measurement values.
5. Statistical determination of accuracy (R).