APG20FC - AZURA Pump P 4.1S with 50mL pump head (hastelloy C), pressure sensor (titanium)
General information
Max. flow rate: 50 ml/min
Flow rate range: 0.01 - 50 ml/min
Flow rate increment: 0.01 ml/min
Maximum delivery pressure: [psi] 2180 psi
Maximum delivery pressure: [MPa] 15 MPa
Maximum delivery pressure: [bar] 150 bar
Pump head materials: Hastelloy® C-276
Maximum viscosity: 100 cp
Liquid temperature range: 4−60 °C (39.2−140 °F)
Gradient: Isocratic
Leak management: No
Wetted materials: aluminum oxide (Al2O3), FEP, FKM, GFP (graphite fiber reinforced PTFE), Hastelloy® C, KEL-F® (PCTFE), PTFE, ruby, sapphire, Tefzel® (ETFC), Titanium, Zirconium oxide
Pump features: Piston seal wash, Purge valve (manual), Standalone control and programming, Integrated pressure sensor
Note: Pressure sensor membrane made of stainless steel; limited wetting in fluidic path
Detailed information
Best working conditions: 1 - 40 ml/min
Continuous working conditions: 3 – 25 ml/min
Flow rate accuracy: ± 2%
Flow rate accuracy conditions: measured at 5 - 50% of flow range using ethanol/water 10:90
Flow rate precision: < 0.5 % RSD
Flow rate precision conditions: measured at 1 mL/min using ethanol/water 10:90
Pump type: Ultra-compact high pressure pump
Pulsation compenation: No
Active piston seal wash: No
Piston seal washing: Yes
System protection: * Pmin and Pmax adjustable * Imin and Imax adjustable
Pump Outlet thread: Female UNF 10-32 (coned)
Pump head inlet (standard): UNF 1/4-28 Thread
Pump head outlet (standard): UNF 10-32 Thread
Communication
Interfaces: Display
Control: RS-232; analog; button on device
Analog inputs: 0 - 10 V
General
Power supply: 100 - 240 V; 50 - 60 Hz Maximum power consumption 40 Watt
Dimensions: 121 mm x 129 mm x 220 mm (W × H × D)
Weight: 2.3 kg
Leak sensor: No
Ambient conditions: temperature range: 10–40 °C; 50–104 °F air humidity: below 90 % humidity (non condensing)
With its small footprint, the compact dual piston AZURA Pump P2.1S can be used for a wide range of laboratory tasks. A pressure rating of up to 400 bar and chemical resistance to a wide range of eluents make it the perfect choice for LC and dosing applications. Pump heads with maximum flow rates of 10 and 50 ml/min are available.
Moreover, its exchangeable pump head allows adaptation of the pump for delivery of aggressive media and bioinert applications. The pump head can be heated or cooled with optional accessories.
OTHER PUMP MODELS ARE AVAILABLE ON REQUEST - Please contact our sales department to discuss your specific requirements.