Water determination in dimethyl sulfoxide (DMSO) by Honeywell | Greyhound Chromatography

Water determination in dimethyl sulfoxide (DMSO) by Honeywell 

HYDRANAL™ Laboratory Report L 141

DMSO by Honeywell

It is not possible to reproduce water determination in DMSO because it alters the stoichiometry of the Karl Fischer (KF) reaction. This shortcoming becomes even greater as the number of samples increases. We tested the water content in different DMSO sample quantities according to the following procedure. 30 mL of Hydranal-Methanol Rapid, Hydranal-Methanol dry or Hydranal-CompoSolver E were introduced into the titration vessel and titrated to dryness with Hydranal-Composite 5. The sample (or sample plus water) was then added and titrated with Hydranal-Composite 5.

Results are compared in the tables below.

Tables 1 & 2 for DMSO research

Coulometric tests gave comparable findings:

• Water content detected in 1 mL DMSO: 311 ppm
• Water content detected in 20 mL DMSO: 266 ppm

Control conditions with Hydranal-Water Standard 1.0 (added water quantity: 1000 ppm) in the presence of DMSO produced the following results:

• Water content in the presence of 1 mL DMSO: 994 ppm
• Water content in the presence of 20 mL DMSO: 910 ppm


The volumetric and coulometric water content determination tests were carried out with two different samples.The problem was picked up again in a series of subsequent tests. We ascertained that it is also impossible to perform indirect determination in a KF oven. A temperature ramp from 50°C to 250°C showed that the water is released by azeotropic distillation at temperatures between 130°C and 190°C. It is impossible to separate the water from the DMSO. The titration vessel is influenced in the same way as by direct injection of the sample. The error is comparable to the effect described for direct injection.

Conclusion

As a result of the various facts, the recommendation must be made that no more than 1 mL of DMSO should be analyzed. The injection of several samples in a coulometric cell should be avoided. Addition of a 1000 ppm standard shows the current effect on the titration vessel.

CONTACT US   

Tel:        +44 (0) 151 649 4000   

Web:     www.greyhoundchrom.com   

Email:   marketing@greyhoundchrom.com   

FOLLOW US   

Facebook Icon Twitter Icon Instagram Icon LinkedIn Icon    

YOU MAY ALSO BE INTERESTED IN OUR NEWSLETTER   

SIGN UP HERE   

Greyhound Monthly Newsletter   

CATALOGUE DOWNLOADS   

Greyhound Q-Fil catalogue  Greyhound reference standards guide   Trajan GC Supplies catalogue 

 Greyhound Chromatography Q-Range Vials Catalogue Image     Chem Service Catalogue Image  Chem Service Pesticides Catalogue Image

Wellington Laboratories Catalogue Image            PFAS EBook  PFAS Environmental EBook

                Wellington Perfluoroalkyl Compounds

Dandy Vice Brochure  Chem Service Catalogue